TOP > Discontinued Model > Calibration Software Medivisor As displays age, luminance and color temperature are apt to change. JVCKENWOOD's Medivisor will keep your displays performing to DICOM standards. Medivisor is included in the calibration kits, along with photo sensor or "puck" for luminance testing and adjustment. FeaturesNew calibration Software – Medivisor NxThe new automatic display detection analyzes display configuration without any wizards or manual inputs. The automatic sensor detection checks for connected and available calibration sensors and selects the right one automatically. Simple user interface makes it easy for you to calibrate the display with just two clicks, no setup or settings necessary. Also one of the new key features is the new tray icon. It shows real-time status of the displays in the Windows task tray. This ensures diagnostic confidence and constant image quality without touching the display. Calibration to conform to DICOM GSDFIt goes without saying that maintaining grayscale accuracy is of significant importance for monochrome medical image displays. Periodic calibration by Medivisor Grayscale and/or Medivisor Color provides accuracy control that conforms to the Grayscale Standard Display Function (GSDF) according to DICOM Part 14. Graphical presentation of calibration resultsCalibration results are shown not only numerically but also graphically providing quick and easy visual summaries. Calibration history managementCalibration history is saved separately for each display and instant access is provided to previous calibration records. User-frendly calibrationAttaching a calibration sensor, luminance,gamma, and color temperature (for colormodels
only) can be simply but accurately calibrated. All that's required is to
follow Medivisor's instructions. In addition, a built-in front sensor provided
with i2 models enables even simpler calibration without attaching the calibration
sensor. *Color temperature calibration by the front sensor is available only for limited models. *Periodic calibration of the front sensor is recommended. It can be performed automatically by calibration using a calibration sensor. Test patterns for easy visual examinationVarious test patterns are included with QA Medivisor, for simple visual examination of display characteristics. QA Medivisor can be used to perform acceptance and constancy tests according to various standards. Supported display models
Supported operating systems
Calibration kit CAL012
|